Section 7: TSP command reference
Series 2600B System SourceMeter® Instrument Reference Manual
7-42
2600BS-901-01 Rev. C / August 2016
Figure 127: ConfigPulseIMeasureVSweepLog()
Example
ConfigPulseIMeasureVSweepLog(smua, 0,
1e-3, 0.01, 1, 1e-3, 10e-3, 10,
smua.nvbuffer1, 5)
Set up a pulsed logarithmic sweep that uses System
SourceMeter
®
instrument channel A. The pulsed
sweep will start at 1 mA, end at 10 mA, and return to
a 0 A bias level between pulses. Each pulsed step
will be on for 1 ms, and then at the bias level for 10
ms. The voltage limit will be 1 V during the entire
pulsed sweep. The pulse train will be comprised of 10
pulsed steps, and the pulse train will be assigned a
tag
index of 5.
Also see
(on page 7-114)
(on page 5-22)
ConfigPulseVMeasureI()
This KIPulse factory script function configures a voltage pulse train with a current measurement at each point.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
No