This read-only buffer recall attribute is like an array (a Lua table) of the status values for all the
readings in the buffer. The status values are floating-point numbers that encode the status value; see
the following table for values.
For dedicated reading buffers, all buffer attributes are saved to nonvolatile memory only when the
reading buffer is saved to nonvolatile memory.
Buffer status bits
Bit
Name
Hex value Description
B1
Overtemp
0x02
Over temperature condition
B2
AutoRangeMeas
0x04
Measure range was autoranged
B3
AutoRangeSrc
0x08
Source range was autoranged
B4
4Wire
0x10
4-wire (remote) sense mode enabled
B5
Rel
0x20
Relative offset applied to reading
B6
Compliance
0x40
Source function was limited because
the complementary function would be
over the compliance limit
B7
Filtered
0x80
Reading was filtered
Also see
(on page 7-25)
(on page 7-26)
(on page 7-27)
(on page 7-28)
(on page 7-29)
(on page 7-31)
(on page 7-32)
bufferVar.timestampresolution
This attribute contains the timestamp's resolution.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Not applicable
See
Details
1e-6 (1
µ
s)
Usage
resolution
=
bufferVar
.timestampresolution
resolution
Timestamp resolution in seconds
bufferVar
The reading buffer; can be a dynamically allocated user-defined buffer or a
dedicated reading buffer