Section 5: Leakage current and insulation resistance
Model 2470 High Voltage SourceMeter Instrument User's Manual
5-10
2470-900-01 Rev. A /
May
2019
Set up the insulation resistance application using the front panel
To set up the application from the front panel:
1.
Connect the device under test (DUT) to the rear panel of the 2470, as described in
(on page 5-2).
2.
Reset the 2470.
a.
Press the
MENU
key.
b.
Under System, select
Info/Manage
.
c.
Select
System Reset
, and then select
OK
.
3.
Press the
TERMINALS FRONT/REAR
switch to set the instrument to use the rear-panel
terminals (R is displayed to the left of the switch).
4.
Press the
HOME
key.
5.
Press the
FUNCTION
key.
6.
Under Source Voltage and Measure, select
Resistance
.
7.
Select the button next to Source (at the bottom of the screen).
8.
Enter
700 V
and select
OK
.
9.
Press the
MENU
key.
10. Under Trigger, select
Templates
.
11. Set Templates to
SimpleLoop
.
12. Set Count to
10
.
13. Set Delay to
0.1
seconds.
14. Press the
HOME
key.
15. Press the
OUTPUT ON/OFF
switch to enable the output.
16. Press the
TRIGGER
key to begin taking readings. The resistance measurements appear in the
measure area (top half) of the Home screen.
17. Press the
OUTPUT ON/OFF
switch to turn off the output when you are done making
measurements.