2-22
Calibration
Table 2-8
Voltage calibration initialization commands
Command
Description
*RST
:SOUR:FUNC VOLT
:SENS:CURR:PROT 0.1
:SENS:CURR:RANG 0.1
:SOUR:VOLT:PROT MAX
:SYST:RSEN OFF
:CAL:PROT:CODE 'KI002425'
:OUTP:STAT ON
Restore GPIB defaults.
Activate voltage source.
Current limit when voltage source is active.
Make sure 1A range is not active.
Maximum allowable source voltage.
Disable remote sensing.
Unlock cal.
Turn source on.
Step 3: Current calibration.
1. Connect the Model 2425 to the digital multimeter (Figure 2-2), and select the mul-
timeter DC current function.
2. Send the commands summarized in Table 2-10 in the order listed to initialize cur-
rent calibration.