Performance Verification
1-15
4. Verify output current accuracy for the 3A range. Be sure to:
•
Select the 3A source range.
•
Set the Model 2425 output current to the correct 3A output value.
•
Verify that the multimeter reading is within the 3A range limits given in Table 1-5.
(Since the value of the 1
Ω
resistor value is assumed to be the same as its nominal
value, the DMM voltage reading is the same as the sourced current.)
5. Repeat the procedure for a negative 3A current output value.
6. Repeat the entire procedure using the rear panel INPUT/OUTPUT jacks. Be sure to
select the rear panel jacks with the front panel TERMINALS key.
Table 1-5
Output current accuracy limits
Model 2425
source range
Model 2425
output current setting
Output current limits
(1 year, 18°C to 28°C)
10mA
100mA
1mA
10mA
100mA
1A
3A
1
10.0000mA
100.000mA
1.00000mA
10.0000mA
100.000mA
1.00000A
3.00000A
9.9947 to 10.0053mA
99.949 to 100.051mA
0.99946 to 1.00054mA
9.9935 to 10.0065mA
99.914 to 100.086mA
0.99893 to 1.00107A
2.99543 to 3.00457A
1
See separate procedure for 3A range. DMM voltage reading is same as sourced current.
Summary of Contents for SourceMeter 2425
Page 3: ...TAIWAN 1FL 85 Po Ai Street Hsinchu Taiwan R O C 886 3 572 9077 Fax 886 3 572 9031...
Page 15: ......
Page 16: ...1 PerformanceVerification...
Page 35: ...1 20 Performance Verification Model 2425 Service Manual...
Page 36: ...2 Calibration...
Page 62: ...3 RoutineMaintenance...
Page 65: ...3 4 Routine Maintenance Model 2425 Service Manual...
Page 66: ...4 Troubleshooting...
Page 84: ...5 Disassembly...
Page 92: ...6 ReplacementParts...
Page 113: ...6 22 Replacement Parts Model 2425 Service Manual...
Page 114: ...A Specifications...
Page 116: ...B CommandReference...
Page 128: ...C CalibrationPrograms...
Page 139: ......
Page 141: ......