Command Reference
B-9
:DATA?
(:CALibration:PROTected:SENSe:DATA?)
Purpose
To request the calibration constants for the active range.
Format
:cal:prot:sens:data?
:cal:prot:sour:data?
Response
Four comma-separated ASCII floating-point constants
Description
The :CAL:PROT:SENS:DATA? and :CAL:PROT:SOUR:DATA? queries
request the calibration constants for the active range of the sense and source
functions, respectively. The four returned constants are in ASCII floating-
point format delimited by commas.
Note
To request the appropriate constants, (1) select the source or sense function,
(2) choose the desired range, and (3) send the desired :DATA? query.
Example
:SENS:FUNC 'VOLT:DC'
Select DC sense function.
:SENS:VOLT:DC:RANGE 0.2
Choose 200mV range.
:CAL:PROT:SENS:DATA?
Request cal constants.
Detecting calibration errors
If an error occurs during any calibration step, the Model 2425 will generate an appropriate
error message. Several methods to detect calibration errors are discussed below.
Reading the error queue
As with other Model 2425 errors, any calibration errors will be reported in the error queue.
(You can read the error queue by using the :SYST:ERR? query.)
Error summary
Table B-4 summarizes calibration errors. These errors will occur under the following
conditions.
Error 500/501
— These errors will occur if you do not set the calibration date or
calibration due date.
Error 502
— This error will occur when the :CAL:PROT:SENS or :CAL:PROT:SOUR
command is sent with a parameter that cannot be placed into any of the defined numeric
ranges.
Summary of Contents for SourceMeter 2425
Page 3: ...TAIWAN 1FL 85 Po Ai Street Hsinchu Taiwan R O C 886 3 572 9077 Fax 886 3 572 9031...
Page 15: ......
Page 16: ...1 PerformanceVerification...
Page 35: ...1 20 Performance Verification Model 2425 Service Manual...
Page 36: ...2 Calibration...
Page 62: ...3 RoutineMaintenance...
Page 65: ...3 4 Routine Maintenance Model 2425 Service Manual...
Page 66: ...4 Troubleshooting...
Page 84: ...5 Disassembly...
Page 92: ...6 ReplacementParts...
Page 113: ...6 22 Replacement Parts Model 2425 Service Manual...
Page 114: ...A Specifications...
Page 116: ...B CommandReference...
Page 128: ...C CalibrationPrograms...
Page 139: ......
Page 141: ......