2400-902-01 (G - Feb 2006)
Model 2400 Service Manual
5-4
Disassembly
Case cover removal
Follow the steps below to remove the case cover to gain access to internal parts.
WARNING
Before removing the case cover, disconnect the line cord and any test leads
from the instrument.
1.
Remove Handle — The handle serves as an adjustable tilt-bail. Adjust its position by
gently pulling it away from the sides of the instrument case and swinging it up or down.
To remove the handle, swing the handle below the bottom surface of the case and back
until the orientation arrows on the handles line up with the orientation arrows on the
mounting ears. With the arrows lined up, pull the ends of the handle away from the
case.
2.
Remove Mounting Ears — Remove the screw that secures each mounting ear. Pull
down and out on each mounting ear.
NOTE
When reinstalling the mounting ears, be sure to mount the right ear to the right side
of the chassis and the left ear to the left side of the chassis. Each ear is marked
“RIGHT” or “LEFT” on its inside surface.
3.
Remove Rear Bezel — To remove the rear bezel, loosen the two screws that secure the
rear bezel to the chassis, and then pull the bezel away from the case.
4.
Removing Grounding Screws — Remove the two grounding screws that secure the
case to the chassis. They are located on the bottom of the case at the back.
5.
Remove Chassis — To remove the case, grasp the front bezel of the instrument, and
carefully slide the chassis forward. Slide the chassis out of the metal case.
NOTE
To gain access to the components under the analog board shield, remove the shield,
which is secured to the analog board by a single screw.
Analog board removal
Perform the following steps to remove the analog board. This procedure assumes that the case
cover is already removed.
1.
Remove analog board shield.
Remove the screw that secures the shield to the analog board, then remove the shield.
2.
Disconnect the front and rear input terminals.
You must disconnect these input terminal connections for both the front and rear inputs:
•
INPUT/OUTPUT HI and LO
•
4-WIRE SENSE HI and LO
•
V,
Ω
, GUARD, and GUARD SENSE (rear panel only)
Summary of Contents for SourceMeter 2400
Page 10: ......
Page 13: ...1 PerformanceVerification...
Page 28: ...2400 902 01 G Feb 2006 Model 2400 Service Manual 1 16 Performance Verification...
Page 29: ...2 Calibration...
Page 51: ...3 RoutineMaintenance...
Page 54: ...2400 902 01 G Feb 2006 Model 2400 Service Manual 3 4 Routine Maintenance...
Page 55: ...4 Troubleshooting...
Page 68: ...2400 902 01 G Feb 2006 Model 2400 Service Manual 4 14 Troubleshooting...
Page 69: ...5 Disassembly...
Page 76: ...2400 902 01 G Feb 2006 Model 2400 Service Manual 5 8 Disassembly...
Page 77: ...6 ReplaceableParts...
Page 91: ...A Specifications...
Page 111: ...B CommandReference...
Page 123: ...C CalibrationPrograms...
Page 130: ...C 8 Calibration Programs...
Page 134: ......
Page 136: ......