Troubleshooting
2400-902-01 (G - Feb 2006)
There are four voltage ranges: 0.2V, 2V, 20V, and 200V. The feedback gain changes for only
the 20V and 200V ranges, resulting in three unique feedback gain values. A multiplexer directs
the voltage feedback, current feedback, reference, or ground signal to the A/D converter. An
opto-isolated interface provides control signals for both DACs, analog circuit control, and A/D
converter communication to the digital section.
V DAC
I DAC
Control
FB
V Clamp
I Clamp
IFB
VFB
-36
+220
Error
Amp
Output
Stage
Remote
VFB
IFB
+7
A/D
MUX
S
Sense
Resistors
Protection
Protection
Output
HI
Output
LO
Guard
Out
Guard
Sense
S+
S-
O
O
O
-220
+36
+
-
Figure 4-1
Analog circuit block diagram
Summary of Contents for SourceMeter 2400
Page 10: ......
Page 13: ...1 PerformanceVerification...
Page 28: ...2400 902 01 G Feb 2006 Model 2400 Service Manual 1 16 Performance Verification...
Page 29: ...2 Calibration...
Page 51: ...3 RoutineMaintenance...
Page 54: ...2400 902 01 G Feb 2006 Model 2400 Service Manual 3 4 Routine Maintenance...
Page 55: ...4 Troubleshooting...
Page 68: ...2400 902 01 G Feb 2006 Model 2400 Service Manual 4 14 Troubleshooting...
Page 69: ...5 Disassembly...
Page 76: ...2400 902 01 G Feb 2006 Model 2400 Service Manual 5 8 Disassembly...
Page 77: ...6 ReplaceableParts...
Page 91: ...A Specifications...
Page 111: ...B CommandReference...
Page 123: ...C CalibrationPrograms...
Page 130: ...C 8 Calibration Programs...
Page 134: ......
Page 136: ......