Model 2400 Service Manual
Calibration
2400-902-01 (G - Feb 2006)
•
Note the multimeter reading.
Table 2-11
Current range calibration commands
Step Command/procedure*
Description
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
:SOUR:CURR:RANGE <Range>
:SOUR:CURR -<Range>
Take DMM reading.
:CAL:PROT:SOUR <DMM_Reading>
Check 2400 for errors.
:CAL:PROT:SENS <DMM_Reading>
Check 2400 for errors.
:SOUR:CURR 0.0
Take DMM reading.
:CAL:PROT:SOUR <DMM_Reading>
Check 2400 for errors.
:CAL:PROT:SENS <DMM_Reading>
Check 2400 for errors.
:SOUR:CURR +<Range>
Take DMM reading.
:CAL:PROT:SOUR <DMM_Reading>
Check 2400 for errors.
:CAL:PROT:SENS <DMM_Reading>
Check 2400 for errors.
:SOUR:CURR 0.0
Take DMM reading.
:CAL:PROT:SOUR <DMM_Reading>
Select source range.
Establish negative polarity.
Read actual output value.
Calibrate sense function negative full scale.
Calibrate source function negative full scale.
Set output to 0A.
Read actual output value.
Calibrate sense function negative zero.
Calibrate source function negative zero.
Establish positive polarity.
Read actual output value.
Calibrate sense function positive full scale.
Calibrate source function positive full scale.
Set output to 0A.
Read actual output value.
Calibrate source positive zero.
*1. Perform complete procedure for each range, where <Range> = 1E6, 10E6, 100E6, 1E3, 10E3, 100E3, or 1.
*
2. <DMM_Reading> parameter is multimeter reading from previous step.
*
3. Use :SYST:ERR? query to check for errors.
•
Send the source and sense calibration commands using the multimeter reading for
the parameter. For example:
:CAL:PROT:SOUR 1E-6
:CAL:PROT:SENS 1E-6
•
Set the source to the positive full-range value using the :SOUR:CURR command.
For example, for the 1mA range:
:SOUR:CURR 1E3
•
Note and record the multimeter reading.
•
Send the source and sense commands using the multimeter reading as the
parameter. For example:
:CAL:PROT:SOUR 1.03E-3
:CAL:PROT:SENS 1.03E-3
•
Send the :SOUR:CURR 0.0 command to set the source current to 0A.
Summary of Contents for SourceMeter 2400
Page 10: ......
Page 13: ...1 PerformanceVerification...
Page 28: ...2400 902 01 G Feb 2006 Model 2400 Service Manual 1 16 Performance Verification...
Page 29: ...2 Calibration...
Page 51: ...3 RoutineMaintenance...
Page 54: ...2400 902 01 G Feb 2006 Model 2400 Service Manual 3 4 Routine Maintenance...
Page 55: ...4 Troubleshooting...
Page 68: ...2400 902 01 G Feb 2006 Model 2400 Service Manual 4 14 Troubleshooting...
Page 69: ...5 Disassembly...
Page 76: ...2400 902 01 G Feb 2006 Model 2400 Service Manual 5 8 Disassembly...
Page 77: ...6 ReplaceableParts...
Page 91: ...A Specifications...
Page 111: ...B CommandReference...
Page 123: ...C CalibrationPrograms...
Page 130: ...C 8 Calibration Programs...
Page 134: ......
Page 136: ......