Table B-3
:CALibration:PROTected:SOURce parameter ranges
Source range
First parameter
(negative full scale)
Second parameter
(negative zero)
Third parameter
(positive full scale)
Fourth parameter
(positive zero)
0.2V
2V
20V
200V
1µA
10µA
100µA
1mA
10mA
100mA
1A
-0.18 to -0.22
-1.8
to -2.2
-18
to -22
-180 to -220
-0.9E-6 to -1.1E-6
-9E-6
to -11E-6
-90E-6 to -110E-6
-0.9E-3 to -1.1E-3
-9E-3
to -11E-3
-90E-3 to -110E-3
-0.9
to -1.1
-0.002 to +0.002
-0.02
to +0.02
-0.2
to +0.2
-2
to +2
-1E-8 to +1E-8
-1E-7 to +1E-7
-1E-6 to +1E-6
-1E-5 to +1E-5
-1E-4 to +1E-4
-1E-3 to +1E-3
-1E-2 to +1E-2
+0.18 to +0.22
+1.8
to +2.2
+18
to +22
+180 to +220
+0.9E-6 to +1.1E-6
+9E-6
to +11E-6
+90E-6 to +110E-6
+0.9E-3 to +1.1E-3
+9E-3
to +11E-3
+90E-3 to +110E-3
+0.9
to +1.1
-0.002 to +0.002
-0.02
to +0.02
-0.2
to +0.2
-2
to +2
-1E-8 to +1E-8
-1E-7 to +1E-7
-1E-6 to +1E-6
-1E-5 to +1E-5
-1E-4 to +1E-4
-1E-3 to +1E-3
-1E-2 to +1E-2
B-8
Command Reference
:DATA?
(:CALibration:PROTected:SENSe:DATA?)
(:CALibration:PROTected:SOURce:DATA?)
Purpose
To request the calibration constants for the active range.
Format
:cal:prot:sens:data?
:cal:prot:sour:data?
Response
Four comma-separated ASCII floating-point constants.
Description
The :CAL:PROT:SENS:DATA? and :CAL:PROT:SOUR:DATA? queries
request the calibration constants for the active range of the sense and source
functions respectively. The four returned constants are in ASCII floating-
point format delimited by commas.
Note
To request the appropriate constants (1) select the source or sense function,
(2) choose the desired range, and (3) send the desired :DATA? query.
Example
:SENS:FUNC ‘VOLT:DC’
Select DC sense function.
:SENS:VOLT:DC:RANGE 0.2
Choose 200mV range.
:CAL:PROT:SENS:DATA?
Request cal constants.
Summary of Contents for SourceMeter 2400
Page 10: ......
Page 13: ...1 PerformanceVerification...
Page 28: ...2400 902 01 G Feb 2006 Model 2400 Service Manual 1 16 Performance Verification...
Page 29: ...2 Calibration...
Page 51: ...3 RoutineMaintenance...
Page 54: ...2400 902 01 G Feb 2006 Model 2400 Service Manual 3 4 Routine Maintenance...
Page 55: ...4 Troubleshooting...
Page 68: ...2400 902 01 G Feb 2006 Model 2400 Service Manual 4 14 Troubleshooting...
Page 69: ...5 Disassembly...
Page 76: ...2400 902 01 G Feb 2006 Model 2400 Service Manual 5 8 Disassembly...
Page 77: ...6 ReplaceableParts...
Page 91: ...A Specifications...
Page 111: ...B CommandReference...
Page 123: ...C CalibrationPrograms...
Page 130: ...C 8 Calibration Programs...
Page 134: ......
Page 136: ......