setmode modifier tables
Modifiers that affect the system
Modifier
Value
Comment
KI_BSWEEP_DELAY
<value> in milliseconds
Adjusts the discharge delay time
for a high-voltage sweep.
KI_CHECK_MODE
0
or
1
0
= Off
1
= On
Setting this to
1
turns on LPTLib
command status error checking
for each LPTLib command.
KI_MAX_VOLTAGE
S540 system, 2657A SMUs only:
<value>
Where:
<value>
=
300
to
3000
Sets the maximum voltage for all
2657A SMUs in the S540
system. The default value is
3000 V.
Calling the
devint
command
resets this value to the
MAX_HV
value defined in the
icconfig_<QMO>.ini
file.
This setmode only affects the
2657A SMU; it does nothing on
other models.
KI_SYSTEM_SPEED_MODE
KI_SYSTEM_SPEED_FAST
KI_SYSTEM_SPEED_LEGACY
FAST
mode optimizes the
S530/S540 for speed with little or
no effect on correlation of data
(default).
LEGACY
mode can be used if you
have problems with correlation of
data using the
FAST
mode.
This setting is reset to the mode
defined in the
icconfig_<QMO>.ini
file with
a call to the
devint
command.