Model 2450 Interactive SourceMeter® Instrument Reference Manual
Section 8: TSP command reference
2450-901-01 Rev. B/September 2013
8-199
trigger.model.setblock() — trigger.BLOCK_BRANCH_ONCE
This attribute causes the trigger model to branch to a specified building block the first time it is encountered in the
trigger model.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Restore configuration
Instrument reset
Power cycle
Configuration script Not applicable
Usage
trigger.model.setblock(
blockNumber
, trigger.BLOCK_BRANCH_ONCE,
branchToBlock
)
blockNumber
The sequence of the block in the trigger model
branchToBlock
The block number of the trigger model block to execute when
the trigger model first encounters this block
Details
The branch-once building block branches to a specified block the first time the trigger model encounters the
branch-once block. If it is encountered again, the trigger model ignores the block and continues in the normal
sequence.
The once block is reset when the trigger model reaches the idle state. Therefore, the branch-once block always
executes the first time the trigger model encounters this block.
Example
trigger.model.setblock(2, trigger.BLOCK_BRANCH_ONCE, 4)
When the trigger model reaches block 2, the trigger model goes to block 4 instead of going in the default
sequence of block 3.