Section 6: SCPI command reference
Model 2450 Interactive SourceMeter® Instrument Reference Manual
6-128
2450-901-01 Rev. B/September 2013
:TRIGger:BLOCk:BRANch:ALWays
This command defines a trigger model block that always goes to a specific block.
Type
Affected by
Where saved
Default value
Command only
Recall settings
Instrument reset
Power cycle
Save settings
Not applicable
Usage
:TRIGger:BLOCk:BRANch:ALWays <blockNumber>, <branchToBlock>
<blockNumber>
The sequence of the block in the trigger model
<branchToBlock>
The block number of the trigger model block to execute when the trigger model
reaches this block
Details
When the trigger model reaches a branch-always building block, it goes to the building block set by
<branchToBlock>
.
Example
TRIG:BLOC:BRAN:ALW 9, 20
When the trigger model reaches block 9, it
will always branch to block 20.
Also see
None
:TRIGger:BLOCk:BRANch:COUNter
This command branches to a specified block a specified number of times.
Type
Affected by
Where saved
Default value
Command only
Recall settings
Instrument reset
Power cycle
Save settings
Not applicable
Usage
:TRIGger:BLOCk:BRANch:COUNter <blockNumber>, <targetCount>, <branchToBlock>
<blockNumber>
The sequence of the block in the trigger model
<targetCount>
The number of times to repeat
<branchToBlock>
The block number of the trigger model block to execute when the counter
is less than to the
<targetCount>
value