ROUTe subsystem
The ROUTe subsystem selects which set of input and output terminals to enable (front panel or rear
panel).
:ROUTe:TERMinals
This command determines which set of input and output terminals the instrument is using.
Type
Affected by
Where saved
Default value
Command and query
Recall settings
Instrument reset
Power cycle
Measure configuration list
Save settings
Measure configuration list
FRON
Usage
:ROUTe:TERMinals <location>
:ROUTe:TERMinals?
<location>
Measure using the front-panel input and output terminals:
FRONt
Measure using the rear-panel input and output terminals:
REAR
Details
This command selects which set of input and output terminals the instrument uses. You can select front panel or
rear panel terminals.
If the output is turned on when you change from one set of terminals to the other, the output is turned off.
Example
:ROUT:TERM REAR
:ROUT:TERM?
Set the instrument to use the rear-panel
terminals and query to verify.
Output:
REAR
Also see
None