Section 3: Using a remote interface
Model DMM7510 7½ Digit Multimeter User's Manual
3-12
DMM7510-900-01 Rev. B / May 2015
Connect to the instrument web interface
When the LAN and instrument establish a connection, you can open a web page for the instrument.
To access the web interface:
1.
Open a web browser on the host computer.
2.
Enter the IP address of the instrument in the address box of the web browser. For example, if the
instrument IP address is 192.168.1.101, enter
192.168.1.101
in the browser address box.
3.
Press
Enter
on the computer keyboard to open the instrument web page.
4.
If prompted, enter a user name and password. The default is
admin
for both.
LAN troubleshooting suggestions
If you are unable to connect to the web interface of the instrument, check the following items:
•
The network cable is in the LAN port on the rear panel of the instrument, not one of the
TSP-Link
®
ports.
•
The network cable is in the correct port on the computer. The LAN port of a laptop may be
disabled when the laptop is in a docking station.
•
The setup procedure used the configuration information for the correct ethernet card.
•
The network card of the computer is enabled.
•
The IP address of the instrument is compatible with the IP address on the computer.
•
The subnet mask address of the instrument is the same as the subnet mask address of the
computer.
You can also try restarting the computer and the instrument. To restart the instrument:
1.
Turn the instrument's power off, and then on.
2.
Wait at least 60 seconds for the network configuration to be completed.
3.
Press the
MENU
key.
4.
Under System, select
Communication
.
5.
Select the
LAN
tab.
6.
Verify the settings.
If the above actions do not correct the problem, contact your system administrator.