Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Section 8: TSP command reference
DMM7510-901-01 Rev. B / May 2015
8-225
Details
When this attribute is enabled, the IEEE Std 488.2 common commands generate prompts if
prompting is enabled with the
localnode.prompts
attribute. If
localnode.prompts4882
is
enabled, limit the number of
*trg
commands sent to a running script to 50 regardless of the setting
of the
localnode.prompts
attribute.
When this attribute is disabled, IEEE Std 488.2 common commands will not generate
prompts. When using the *trg command with a script that executes trigger.wait()
repeatedly, disable prompting to avoid problems associated with the command
interface input queue filling.
Example
localnode.prompts4882 = localnode.DISABLE
Disables IEEE Std 488.2 common command
prompting.
Also see
(on page 8-223)
localnode.serialno
This attribute stores the instrument's serial number.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (R)
Yes
Not applicable
Not applicable
Not applicable
Usage
serialno
= localnode.serialno
serialno
The serial number of the instrument
Details
This indicates the instrument serial number.
Example
display.clear()
display.settext(display.TEXT2, "Serial #: " ..localnode.serialno)
display.changescreen(display.SCREEN_USER_SWIPE)
Clears the instrument display.
Places the serial number of this instrument on the bottom line of the USER swipe screen display. Displays the
USER swipe screen.
Also see
(on page 8-222)