Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Section 8: TSP command reference
DMM7510-901-01 Rev. B / May 2015
8-223
localnode.password
This attribute stores the instrument password.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (W)
No
Rear-panel LAN reset
Nonvolatile memory
"admin"
Usage
localnode.password = "
password"
password
A string that contains the instrument password (maximum 30 characters)
Details
When the access to the instrument is set to protected or lockout, this is the password that is used to
gain access.
If you forget the password, you can reset the password to the default:
1. On the front panel, press
MENU
.
2. Under System, select
Info/Manage
.
3. Select
Password Reset
.
You can also reset the password and the LAN settings from the rear panel by inserting a
straightened paper clip into hole below LAN RESET.
Example
localnode.password = "N3wpa55w0rd"
Changes the password to
N3wpa55w0rd
.
Also see
(on page 8-220)
localnode.prompts
This attribute determines if the instrument generates prompts in response to command messages.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
No
Power cycle
Not saved
localnode.DISABLE
Usage
prompting
= localnode.prompts
localnode.prompts =
prompting
prompting
Do not generate prompts:
localnode.DISABLE
Generate prompts:
localnode.ENABLE