Section 8: TSP command reference
Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
8-192
DMM7510-901-01 Rev. B / May 2015
dmm.measure.thermocouple
This attribute indicates the thermocouple type.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Restore configuration
Instrument reset
Power cycle
Measure configuration list
Configuration script
Measure configuration list
dmm.THERMOCOUPLE_K
Usage
value
= dmm.measure.thermocouple
dmm.measure.thermocouple =
value
value
The thermocouple type:
•
dmm.THERMOCOUPLE_B
•
dmm.THERMOCOUPLE_E
•
dmm.THERMOCOUPLE_J
•
dmm.THERMOCOUPLE_K
•
dmm.THERMOCOUPLE_N
•
dmm.THERMOCOUPLE_R
•
dmm.THERMOCOUPLE_S
•
dmm.THERMOCOUPLE_T
Functions
dmm.FUNC_DC_VOLTAGE
dmm.FUNC_RESISTANCE
dmm.FUNC_ACV_FREQUENCY
dmm.FUNC_AC_VOLTAGE
dmm.FUNC_4W_RESISTANCE
dmm.FUNC_ACV_PERIOD
dmm.FUNC_DC_CURRENT
dmm.FUNC_DIODE
dmm.FUNC_DCV_RATIO
dmm.FUNC_AC_CURRENT
dmm.FUNC_CAPACITANCE
dmm.FUNC_DIGITIZE_CURRENT
dmm.FUNC_TEMPERATURE
dmm.FUNC_CONTINUITY
dmm.FUNC_DIGITIZE_VOLTAGE
Details
This command is only applicable when the transducer type is set to thermocouple.
Example
dmm.measure.func = dmm.FUNC_TEMPERATURE
dmm.measure.transducer = dmm.TRANS_THERMOCOUPLE
dmm.measure.thermocouple = dmm.THERMOCOUPLE_J
Set the measure function to temperature.
Set the transducer type to thermocouple.
Set the thermocouple type to J.
Also see
(on page 8-190)
(on page 2-120)