Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
Section 8: TSP command reference
DMM7510-901-01 Rev. B / May 2015
8-175
dmm.measure.range
This attribute determines the positive full-scale measure range.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Restore configuration
Instrument reset
Power cycle
Measure configuration list
Configuration script
Measure configuration list
See
Details
Usage
rangeValue
= dmm.measure.range
dmm.measure.range =
rangeValue
rangeValue
See
Details
Functions
dmm.FUNC_DC_VOLTAGE
dmm.FUNC_RESISTANCE
dmm.FUNC_ACV_FREQUENCY
dmm.FUNC_AC_VOLTAGE
dmm.FUNC_4W_RESISTANCE
dmm.FUNC_ACV_PERIOD
dmm.FUNC_DC_CURRENT
dmm.FUNC_DIODE
dmm.FUNC_DCV_RATIO
dmm.FUNC_AC_CURRENT
dmm.FUNC_CAPACITANCE
dmm.FUNC_DIGITIZE_CURRENT
dmm.FUNC_TEMPERATURE
dmm.FUNC_CONTINUITY
dmm.FUNC_DIGITIZE_VOLTAGE
Details
You can assign any real number using this command. The instrument selects the closest fixed range
that is large enough to measure the entered number. For example, for current measurements, if you
expect a reading of approximately 9 mA, set the range to 9 mA to select the 10 mA range. When you
read this setting, you see the positive full-scale value of the measurement range that the instrument is
presently using.
This command is primarily intended to eliminate the time that is required by the instrument to
automatically search for a range.
When a range is fixed, any signal greater than the entered range generates an overrange condition.
When an overrange condition occurs, the front panel displays "Overflow" and the remote interface
returns
9.9e+37
.
When you set a value for the measurement range, the measurement autorange setting is
automatically disabled for the selected measurement function.
The range for measure functions defaults to autorange. When you switch from autorange to range,
the range is set to the last selected autorange value.