Section 8: TSP command reference
Model DMM7510 7½ Digit Graphical Sampling Multimeter Reference Manual
8-122
DMM7510-901-01 Rev. B / May 2015
Example
dmm.measure.func = dmm.FUNC_DC_CURRENT
dmm.measure.range = 3
dmm.measure.autozero.enable = dmm.OFF
dmm.measure.analogtrigger.mode = dmm.MODE_EDGE
dmm.measure.analogtrigger.edge.level = 5
dmm.measure.analogtrigger.edge.slope = dmm.SLOPE_FALLING
Set measure function to DC current.
Set range to 3 A.
Disable autozero.
Set the analog trigger mode to edge.
Set the analog trigger level to 2.5 A.
Set the level to be detected on a falling edge.
Also see
Analog triggering example with digitize function
(on page 3-67)
(on page 3-64)
dmm.digitize.analogtrigger.mode
(on page 8-74)
dmm.measure.analogtrigger.pulse.condition
This attribute defines if the pulse must be greater than or less than the pulse width before an analog trigger is
generated.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
Restore configuration
Instrument reset
Power cycle
Measure configuration list
Configuration script
Measure configuration list
dmm.CONDITION_GREATER
Usage
value
= dmm.measure.analogtrigger.pulse.condition
dmm.measure.analogtrigger.pulse.condition =
value
value
The setting:
•
The pulse width must be greater than the specified pulse width:
dmm.CONDITION_GREATER
•
The pulse width must be less than the specified pulse width:
dmm.CONDITION_LESS
Functions
dmm.FUNC_DC_VOLTAGE
dmm.FUNC_RESISTANCE
dmm.FUNC_ACV_FREQUENCY
dmm.FUNC_AC_VOLTAGE
dmm.FUNC_4W_RESISTANCE
dmm.FUNC_ACV_PERIOD
dmm.FUNC_DC_CURRENT
dmm.FUNC_DIODE
dmm.FUNC_DCV_RATIO
dmm.FUNC_AC_CURRENT
dmm.FUNC_CAPACITANCE
dmm.FUNC_DIGITIZE_CURRENT
dmm.FUNC_TEMPERATURE
dmm.FUNC_CONTINUITY
dmm.FUNC_DIGITIZE_VOLTAGE
Details
Only available when the analog trigger mode is set to pulse.