Section 4: TSP command reference
Model DMM7510 7½ Digit Graphical Sampling Multimeter Calibration Manual
4-4
DMM7510-905-01 Rev. A / April 2015
acal.lastrun.time
This attribute returns the date and time when auto calibration was last run.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (R)
Yes
Not applicable
Not applicable
Not applicable
Usage
dateTime
= acal.lastrun.time
dateTime
The date and time
Details
The date and time is returned in the format:
MM/DD/YYYY HH:MM:SS.NNNNNNNNN
Where:
•
MM/DD/YYYY
is the month, date, and year
•
HH:MM:SS.NNNNNNNNN
is the hour, minute, second, and fractional second
Example
print(acal.lastrun.time)
Returns the date and time when the auto
calibration was last run.
Example output:
05/28/2014 00:37:47.743000000
Also see
(on page 4-6)