Model 8011 Connection Instructions
Specifications are subject to change without notice.
PA-1032 Rev. A / April 2011
11
Table 3. Connections for three-terminal device test
Cable end
Cable
Cable end
Model 2651A rear panel
CA-552-1 (supplied with the Model 2651A)
Open
Model 2651A rear panel
CA-557-1 (supplied with the Model 2651A)
CS-1629-8
Series 2600A rear panel
CA-557-x (recommended; separately purchased)
CS-1629-8
LO of CA-552-1
Black CA-560-0
8010-120 yellow terminal 6
HI of CA-552-1
Red CA-560-2
8010-120 white terminal 5
Sense LO of CS-1629-8
Black CA-567-0
8010-120 yellow terminal 3
Sense HI of CS-1629-8
Red CA-567-2
8010-120 white terminal 2
Sense LO of CS-1629-8
Black CA-567-0
8010-120 yellow terminal 3
LO of CS-1629-8
Black CA-567-0
8010-120 yellow terminal 3
Sense HI of CS-1629-8
Red CA-567-2
8010-120 blue terminal 1
HI of CS-1629-8
Red CA-567-2
8010-120 blue terminal 4
Figure 7. Connections for three-terminal device test
G
D
S
LO
S LO
G
HI
G
G
G
S HI
S LO
G
G
G
G
S HI
HI
2600-KIT
3-T
erminal
DUT
HIGH CURRENT TEST SOCKET KIT
8011
RATINGS:
40V MAX.
15A DC MAX.
50A PULSE MAX.
Model 2651A
CA-567-2
(red)
CA-560-0
(black)
CA-560-2
(red)
CS-1629-8
CA-552-1
(supplied with
Model 2651A)
CA-557-1
(supplied with
Model 2651A)
CA-567-0
(black)
Customer-supplied cable
2600-KIT
Model 2601A, 2611A,
2602A, or 2612A
CA-567-2
(red)
CA-567-0
(black)
CS-1629-8
Model 2651A
Detail of stacked
banana plugs
at terminal 3
8010-120 TO-247
Device Test Board