Section 3: Using the web interface
Models 707B and 708B 199BSwitching Matrix User's Manual
3-6
707B-900-01 Rev. B / June 2013
4. You must log into the instrument to work with the channels. See
3-3). After logging in, you can access the channel controls.
To close channel "1A01", click the channel.
Figure 17: Close a channel
To perform an exclusive close on channel "1A02", select
Exclusive Slot Close
in the Channel
Action Type box.
Figure 18: Select exclusive slot close
Click "1A02" to close that channel and open all other channels.
Figure 19: Close a channel with exclusive close
Set up channel patterns from the web interface
You can use channel patterns as a convenient way to refer to a group of switching channels with a
single alphanumeric name. When you perform close or open operations on a channel pattern, only
the channels that are in the channel pattern are affected.
There is no speed advantage in performing close and open operations on channel patterns compared
to performing the same operations on individual channels or a list of channels.