Models 707B and 708B Switching Matrix Reference Manual
Section 7: Command reference
707B-901-01 Rev. A / August 2010
7-223
tsplink.trigger[N].release()
This function releases a latched trigger on the given TSP-Link trigger line.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function Yes
Usage
tsplink.trigger[N].release()
N
The trigger line (1 to 3)
Details
Releases a trigger that was asserted with an indefinite pulse width, as well as a trigger that was latched in
response to receiving a synchronous mode trigger.
Example
tsplink.trigger[3].release()
Releases trigger Line 3.
Also see
(on page 7-219)
(on page 7-225)
tsplink.trigger[N].reset()
This function resets some of the TSP-Link trigger settings to their factory defaults.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function Yes
Usage
tsplink.trigger[N].reset()
N
The trigger line (1 to 3)
Details
The
tsplink.trigger[N].reset()
function resets the following attributes to their factory defaults:
•
tsplink.trigger[N].mode
•
tsplink.trigger[N].stimulus
•
tsplink.trigger[N].pulsewidth
This also clears
tsplink.trigger[N].overrun
.