5-8
Table 5-2
Path isolation tests
Test
no.
Path isolation
Test equipment locations
Channels closed
1
Row 1, Col 1 to Row 2, Col 2
Row 1 and Row 2
1!1!1 and 1!2!2
2
Row 2, Col 2 to Row 3, Col 3
Row 2 and Row 3
1!2!2 and 1!3!3
3
Row 3, Col 3 to Row 4, Col 4
Row 3 and Row 4
1!3!3 and 1!4!4
4
Row 3, Col 4 to Row 4, Col 5
Row 3 and Row 4
1!3!4 and 1!4!5
5
Row 3, Col 5 to Row 4, Col 6
Row 3 and Row 4
1!3!5 and 1!4!6
6
Row 3, Col 6 to Row 4, Col 7
Row 3 and Row 4
1!3!6 and 1!4!7
7
Row 3, Col 7 to Row 4, Col 8
Row 3 and Row 4
1!3!7 and 1!4!8
8
Row 3, Col 8 to Row 4, Col 9
Row 3 and Row 4
1!3!8 and 1!4!9
9
Row 3, Col 9 to Row 4, Col 10
Row 3 and Row 4
1!3!9 and 1!4!10
Figure 5-5
Path isolation testing (guarded)
1
2
3
4
1
3
4
5
6
7
8
9
10
2
Model 7012
HI
Model 617
INPUT
Model 7025
Unterminated
Triax Cable
(Red)
Banana to Banana Cable
Ground Link
Removed
Unterminated
Banana Cables
Source V and
Measure V/I
Note : Setup shown is configured
to test isolation between
row 1 column 1, and row 2
column 2.
HI
Columns
Jumpers
(1 of 4)
Rows
H L
H L
H L
H L
H L
H L
H L
H L
H L
H L
H
L
H
L
H
L
H
L
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com