2-3
2.3
Typical matrix switching schemes
The following paragraphs describe some basic switch-
ing schemes that are possible with a two-pole switch-
ing matrix. These switching schemes include some
various shielding con
Þ
gurations to help minimize
noise pick up in sensitive measurement applications.
These shields are shown connected to chassis ground.
For some test con
Þ
gurations, shielding may prove to
be more effective connected to circuit common. Chassis
ground is accessible at the rear panel of the Model 7001.
2.3.1 Single-ended switching
In the single-ended switching con
Þ
guration, the source
or measure instrument is connected to the DUT
through a single pathway as shown in Figure 2-4.
Figure 2-4
Single-ended switching example
Source or
Measure
Row
Columns
7012
H
L
HI
LO
DUT
Optional
Shield
Figure 2-3
Matrix row connections to backplane
7001
Analog
Backplane
H
L
G
H
L
7012
Matrix Row
(1 of 4)
Backplane
Jumpers
H = High
L = Low
G = Guard