5-87
:STIMe <n>
[:ROUTe]:CONFigure:SLOT[1]:STIMe <n>
Set delay for Slot 1
[:ROUTe]:CONFigure:SLOT2:STIMe <n>
Set delay for Slot 2
<n> = 0 to 99999.999
Settling time in seconds
= DEFault
0 sec
= MINimum
0 sec
= MAXimum
99999.999 sec
:conf:slot:stim <n>
:conf:slot2:stim <n>
Power-up
0 sec
*RST
0 sec
:SYSTem:PRESet
0 sec
:STIMe?
Query programmed delay for specified slot
:STIMe? DEFault
Query *RST default delay
:STIMe? MINimum
Query minimum delay
:STIME? MAXimum
Query maximum delay
Short-form formats:
:conf:slot:stim?
:conf:slot2:stim?
:conf:slot:stim?
def
:conf:slot2:stim?
def
:conf:slot:stim?
min
:conf:slot2:stim?
min
:conf:slot:stim?
max
:conf:slot2:stim?
max
Response message:
0 to 99999.999
An internally set relay settling time is provided to allow for switch bounce when it closes. At
power-up, a settling time is automatically set based on the acquired identification information
from the card. The :STIMe command allows you to increase the delay for each slot.
The delay value can be entered using the exponent form. For example, instead of sending a pa-
rameter value of 2000 (seconds), you can instead send 2E3. To cancel this delay, send this com-
mand using a parameter value of 0.
For complete information on this delay, see DELAY in Section 4.
10
OUTPUT 707; “:conf:slot:stim 10; stim?”
20
ENTER 707; A$
30
PRINT A$
40
END
Line 10
Sets a 10 second delay for Slot 1 and then queries the delay period.
Line 20
Addresses the 7001 to talk.
Line 30
Displays the delay period (10).
5.13.10
Parameters
Short-form
formats
Defaults
Query
commands
Description
Programming
example
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com
Summary of Contents for 7001
Page 284: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Page 285: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Page 286: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Page 287: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Page 288: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Page 293: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Page 294: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Page 295: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Page 296: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Page 299: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Page 300: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Page 301: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Page 304: ......
Page 305: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Page 308: ......
Page 309: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Page 312: ......
Page 313: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Page 315: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...