5.8
General bus commands
General bus commands are those commands such as DCL
that have the same general meaning regardless of the instru-
ment. Commands supported by the Model 7001 are listed in
Table 5-3 which also lists BASIC statements necessary to
send each command. Note that commands requiring that a
primary address be specified assume that the primary address
is set to 7. This is the address that is set at the factory.
5.8.1 REN (remote enable)
The remote enable command is sent to the Model 7001 by
the controller to set up the instrument for remote operation.
Generally, the instrument should be placed in the remote
mode before you attempt to program it over the bus. Simply
setting REN true does not actually place the instrument in the
remote state. Instead the instrument must be addressed to lis-
ten after setting REN true before it goes into remote.
Note that the instrument need not be in remote to be a talker.
Programming Example
This sequence is automatically
performed by the controller when the following is typed into
the keyboard.
REMOTE 707
After the RETURN key is pressed, the instrument is in the
remote state, as indicated by the REM annunciator. If not,
check to see that the instrument is set to the proper primary
address 07, and check to see that the bus connections are
properly made.
Note that all front panel controls except for LOCAL (and, of
course, POWER) are inoperative while the instrument is in
remote. You can restore normal front panel operation by
pressing the LOCAL key.
Table 5-3
General bus commands and associated BASIC statements
Command
HP BASIC 4.0
statement
Effect on Model 7001
REN
REMOTE 7
Goes into effect when next addressed to listen.
IFC
ABORT 7
Goes into talker and listener idle states.
LLO
LOCAL LOCKOUT 7
LOCAL key locked out.
GTL
LOCAL 707
Cancel remote, restore front panel operation.
DCL
CLEAR 7
Returns 7001 to known conditions.
SDC
CLEAR 707
Returns all devices to known conditions.
GET
TRIGGER 707
Initiates a trigger.
SPE, SPD
SPOLL (707)
Serial Polls the Model 7001.
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com
Summary of Contents for 7001
Page 284: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Page 285: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Page 286: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Page 287: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Page 288: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Page 293: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Page 294: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Page 295: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Page 296: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Page 299: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Page 300: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Page 301: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Page 304: ......
Page 305: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Page 308: ......
Page 309: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Page 312: ......
Page 313: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...
Page 315: ...Artisan Scientific Quality Instrumentation Guaranteed 888 88 SOURCE www artisan scientific com...