Model 6517B Electrometer User’s Manual
Section 4: Measurement Options
Speed
SPEED sets the integration time of the A/D converter, the period of time the input signal is
measured. The SPEED can be set for each measurement function. The integration time is
specified in parameters based on a number of power line cycles (NPLC), where one PLC for 60Hz
is 16.67msec and one PLC for 50Hz and 400Hz is 20msec.
In general, the fastest integration time (0.01 PLC) results in increased reading noise and less
usable resolution, while the slowest integration time (10 PLC) provides the best common-mode
and normal-mode rejection. In-between settings are a compromise between speed and noise.
Configuring speed
Each measurement function can have a unique integration time. Speed is set from the configure
function menu structures. The procedure is summarized as follows:
1.
Press
the
CONFIG
key and then the desired function key.
2.
Use the cursor keys to select
SPEED
and press
ENTER
to display the following speed
options:
NORMAL:
Sets the integration time to 1 PLC.
FAST:
Sets the integration time to 0.01 PLC.
MEDIUM:
Sets the integration time to 0.1 PLC.
HIACCURACY:
Sets the integration time to 10 PLC.
SET-SPEED-EXACTLY:
This options prompts you to enter a PLC value (0.01 to 10)
SET-BY-RESOLUTION:
Automatically optimizes the integration time for the present
resolution setting (see Section 4 of the Reference Manual).
3.
Use the menu items to configure speed. A menu item is selected by placing the cursor
(blinking menu item) on it and pressing
ENTER
. Parameter values are changed using the
cursor keys and the
RANGE
keys, and then pressing
ENTER
information on navigating through the menu structure).
For more information, see Section 4 of the Reference Manual.
Resolution
All functions can operate from 3½ to 6½ digits of resolution. Each function can have a unique
resolution setting.
Configuring resolution
Perform the following steps to set resolution for a measurement function:
1.
Press
the
CONFIG
key and then the desired function key.
2.
Using the cursor keys, select
RESOLUTION
and press
ENTER
to display the following
resolution options:
a.
3.5 to 6.5d:
Select one of these options to set the display resolution.
b.
AUTO:
This option automatically optimizes the resolution for the present integration
time (speed) setting (see Section 5 of the Reference Manual).
3.
Set resolution by placing the cursor on the menu option and pressing
ENTER
.
Filter
Filtering stabilizes noisy measurements. The Model 6517B uses a digital filter and a median filter.
When a filter is enabled by pressing
FILTER
(FILT annunciator turns on), the selected filter for that
measurement function is in effect. Pressing
FILTER
a second time disables the filter.