2-18
Measurement Options
Model 6517A Getting Started Manual
Configuring filters
Each measurement function has its own filter configuration. Perform the following steps to
configure filter:
1.
Select the measurement function.
2.
Press CONFIG and then FILTER to display the following menu items:
AVERAGING:
This item is used to control and configure the digital filter.
MEDIAN:
This item is used to control and configure the median filter.
3.
The menu structure for the two filters (Digital and median) is shown as follows:
a.
AVERAGING (Digital) filter options:
TYPE — This item is used to select the digital filter type:
NONE — Disables the digital filter.
AVERAGING — Selects the averaging filter and prompts you to enter the
number of readings to average (“stack” size).
ADVANCED — Selects the advanced filter and prompts you to enter the
number of readings to average (“stack” size). After entering the stack size,
you are then prompted to enter the noise tolerance level (0 to 100%).
AVERAGING-MODE — This menu item is used to select the filter mode;
MOVING or REPEAT.
b.
MEDIAN filter options:
DISABLE — Disables the median filter.
ENABLE — Enables the median filter and prompts you to enter the range (1
to 5).
4.
Use the menu items to configure a filter. A menu item is selected by placing the
cursor (using
and
keys) on it and pressing ENTER. Parameter values are
changed using the cursor keys and the RANGE keys, and then pressing ENTER.
See “Menus” if you need more information to navigate through the menu structure.
NOTE
The filter configuration menus can also be accessed from the configure function
menus.