Model 6517A Getting Started Manual
Front Panel Operation
1-23
Measurement control
BENCH reset disables temperature and humidity measurements. These measurements can
be enabled from the front panel as follows:
1.
Press MENU to display the main menu.
2.
Use the
and
keys to place the cursor (blinking menu item) on GENERAL and
press ENTER to display the menu items of the general menu.
3.
Place the cursor on A/D-CONTROLS and press ENTER to display the A/D con-
trols menu.
4.
Place the cursor on DATA-STAMP and press ENTER. The present control state for
temperature and humidity will be displayed. OFF indicates that the measurement is
disabled, while ON indicates that the measurement is enabled.
5.
To change the control state for temperature and/or humidity, place the cursor on the
appropriate menu item and use the RANGE
and
keys to toggle the control
state. With the desired states for temperature and humidity displayed (ON or OFF),
press ENTER.
6.
Use the EXIT key to back out of the menu.
NOTE
The TEMPERATURE option also controls the temperature reading for the
“Change From Cal Temp” multiple display. See Section 2, “Multiple Displays.”
Temperature units
All temperature readings by the Model 6517A can be displayed in °C, °F, or K. BENCH
reset selects °C measurement units. Perform the following steps to change units:
1.
Press MENU to display the main menu.
2.
Use the
and
keys to place the cursor (blinking menu item) on GENERAL and
press ENTER to display the menu items of the general menu.
3.
Place the cursor on DISPLAY and press ENTER to display the SET READING
DISPLAY menu.
4.
Place the cursor on TEMP-UNITS and press ENTER. The blinking cursor indi-
cates the present temperature units selection.
5.
To change temperature units, place the cursor on the desired selection (°C, K, or
°F) and press ENTER.
6.
Use the EXIT key to back out of the menu.