Model 6487 Reference Manual
DDC Emulation Commands
C-15
Voltage Source Error (B7)
— Set if the voltage source has reached current limit or if the
interlock is tripped. Cleared by reading U9 status after correcting the voltage source error
(see
).
klqb
Once the Model 6487 has generated an SRQ, its status byte must be read to
clear the SRQ line. Otherwise, the instrument will continuously assert SRQ.
Figure C-4
Status byte format
Table C-2
Status byte and SRQ mask interpretation
Bit
Decimal
Weighting
Description
0 (LSB)
1
Reading Overflow
1
2
Data Store Full
2
4
Data Store 1/2 Full
3
8
Reading done
4
16
Ready
5
32
Error
6
64
RQS (Request for Service)*
7
128
Voltage Source Error
* Status byte only
MSB
B7
B6
B5
B4
B3
B2
B1
B0
LSB
Reading
Done
Data
Store
Full
Reading
Overflow
Error
RQS*
Data
Store
1/2 Full
Ready
V-source
Error
* Status bye only