Model 6487 Reference Manual
Calibration
16-5
Calibration menu
You can access the calibration menu by pressing MENU, selecting CAL, and then press-
ing ENTER. The various selections are summarized in
. Use the up and down
RANGE keys to scroll through these selections.
Table 16-2
Current test uncertainty ratios with recommended equipment
Range
5700 + 5156
Test uncertainty
ratio
2nA
7ppm + 300ppm
13.0
20nA
7ppm + 200ppm
19.3
200nA
5ppm + 200ppm
9.8
2
μ
A
7ppm + 200ppm
7.2
20
μ
A
550ppm
1.8
200
μ
A
100ppm
10.0
2mA
55ppm
18.2
20mA
55ppm
18.2
Table 16-3
Voltage source test uncertainty ratios with recommended equipment
Range
2001 DMM
Test uncertainty
ratio
10V
32ppm
31.2
50V
50ppm
20.0
500V
53ppm
28.3