15
Performance Verification
•
Verification test requirements
— Summarizes environmental conditions, warm-
up period, and line power requirements.
•
— Lists all equipment necessary for verification
and gives pertinent specifications.
•
— Describes how reading limits are calculated and gives an
example.
•
Calibrator voltage calculations
— Details the method for calculating calibrator
voltages when testing the 2nA to 2
μ
A ranges.
•
Performing the verification test procedures
— Summarizes test considerations
and describes how to restore factory defaults.
•
— Lists steps necessary to null voltage offset before
performing verification.
•
— Lists detailed steps for verifying measure-
ment accuracy of all current ranges. The 2nA to 2
μ
A and 20
μ
A to 20mA ranges
are covered separately because of the different test equipment required.
•
Voltage source output accuracy
— Summarizes the method to test voltage source
output accuracy.