14-18
SCPI Reference Tables
Model 6487 Reference Manual
Table 14-9
TRIGger command summary
Command
Description
Default
parameter
Ref
Section SCPI
:INITiate
Path to initiate measurement cycle(s):
√
[:IMMediate]
Initiate one trigger cycle.
√
:ABORt
Reset trigger system (goes to idle state).
√
:ARM[:SEQuence[1]]
Path to configure arm layer:
√
[:LAYer[1]]
:SOURce <name>
Select control source: IMMediate, TIMer,
BUS, TLINk, MANual, PSTest, NSTest, or
BSTest.
IMM
√
:SOURce?
Query arm control source.
√
:COUNt <n>
Set measure count: 1 to 2048 or INF (infinite).
1
1
√
:COUNt?
Query measure count.
√
:TIMer <n>
Set timer interval: 0.001 to 99999.999 (sec).
0.100
√
:TIMer?
Query timer interval.
√
[:TCONfigure]
√
:DIRection <name>
Enable (SOURce) or disable (ACCeptor)
bypass.
ACC
√
:DIRection?
Query arm source bypass.
√
[:ASYNchronous]
Configure input/output triggers:
:ILINe <NRf>
Select input trigger line: 1, 2, 3, 4, 5, or 6.
1
:ILINe?
Query input trigger line.
:OLINe <NRf>
Select output trigger line: 1, 2, 3, 4, 5, or 6.
2
:OLINe?
Query output trigger line.
:OUTPut <name>
Output trigger (TRIGger) or not at all
(NONE).
NONE
:OUTPut?
Query output trigger status.
1.
*RST default is 1. SYST:PRES default is INF.