10-6
Return to
6482-901-01 Rev. A / August 2012
Section 10: Limit Testing
Model 6482 Dual-Channel Picoammeter / Voltage Source Reference Manual
•
S/W-LIMITS
: Use this menu item to control, set limits for, and define output bit patterns for
Limit 3 to Limit 6 tests:
–
CONTROL
: Use to ENABLE or DISABLE the test.
–
LOLIM
: Use to set the low limit.
–
HILIM
: Use to set the high limit.
–
FEED
: Use to select the data source for the Limit 3 through Limit 6 tests. Specify MSR1
or MSR2 (channel 1 or channel 2 measurement function), RATIO, or DELTA.
Figure 10-3
Limits configuration menu tree
Performing limit tests
Perform the following steps to run limit tests:
Step 1. Configure test system.
As previously explained in Section 2, your test system could be as simple as connecting one or
two DUTs to the Model 6482. Adding a scanner to the test system allows you to test multiple
devices.
Step 2. Configure bias source and measure functions.
Configure the Model 6482 operation as follows:
1. Select the source channel by pressing the
SRC1
or
SRC2
key.
2. Set the source level value.
3. Press the
MSR1
or
MSR2
key to select the measurement channel, then choose the measure-
ment range.
Refer to
,
for more information.
Step 3. Configure limit tests.
Select and configure the following limit tests parameters as explained in
:
•
Set the H/W-LIMITS and S/W-LIMITS parameters.
Config
Digout
Mode
Size
Control
Control
Feed
Limits
3-6
H/W
limits 1, 2
S/W
limits
Fail
mode
LOLIM HILIM
Limit
Auto
Clear