Routine Maintenance
20-3
Perform the following steps to replace the line fuse:
1.
Carefully grasp and squeeze together the locking tabs that secure the fuse carrier to the
fuse holder.
2.
Pull out the fuse carrier, and replace the fuse with the type speci
fi
ed in Table 20-1.
CAUTION
To prevent instrument damage, use only the fuse type specified in
Table 20-1.
3.
Reinstall the fuse carrier.
NOTE
If the power line fuse continues to blow, a circuit malfunction exists and must be
corrected. Refer to the troubleshooting section of this manual for additional
information.
Front panel tests
There are three front panel tests: one to test the functionality of the front panel keys and two
to test the display.
KEYS test
The KEYS test lets you check the functionality of each front panel key. Perform the follow-
ing steps to run the KEYS test.
1.
Display the MAIN MENU by pressing the MENU key.
2.
Select TEST, and press ENTER to display the SELF-TEST MENU.
3.
Select DISPLAY-TESTS, and press ENTER to display the following menu:
FRONT PANEL TESTS
KEYS DISPLAY-PATTERNS CHAR-SET
4.
Select KEYS, and press ENTER to start the test. When a key is pressed, the label name
for that key will be displayed to indicate that it is functioning properly. When the key is
released, the message
“
No keys pressed
”
is displayed.
5.
Pressing EXIT tests the EXIT key. However, the second consecutive press of EXIT
aborts the test and returns the instrument to the SELF-TEST MENU. Continue pressing
EXIT to back out of the menu structure.
Table 20-1
Power line fuse
Line voltage
Rating
Keithley part no.
100-240V
250V, 2.5A, Slow
Blow 5
×
20mm
FU-72
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......