Performance Verification
18-9
The
“
real
”
compliance condition can occur when the compliance setting is less than the
highest possible reading of the measurement range. When in compliance, the source output
clamps at the displayed compliance value. For example, if the compliance voltage is set to 1V
and the measurement range is 2V, the output voltage will clamp (limit) at 1V.
“
Range
”
compliance can occur when the compliance setting is higher than the possible
reading of the selected measurement range. When in compliance, the source output clamps at
the maximum measurement range reading (not the compliance value). For example, if the com-
pliance voltage is set to 1V and the measurement range is 200mV, the output voltage will clamp
(limit) at 210mV.
Maximum compliance values
The maximum compliance values for the measurement ranges are summarized in
Table 18-2.
When the unit goes into compliance, the
“
Cmpl
”
label or the units label (i.e.,
“
mA
”
) for the
compliance display will
fl
ash.
Table 18-2
Maximum compliance values
Measurement range
1
Maximum
compliance value
200mV
00
2V
0
20V
200V
00
1pA
0
10pA
100pA
00
1nA
0
10nA
100nA
00
1µA
0
10µA
100µA
00
1mA
0
10mA
100mA
210mV
00
2.1V
0
21V
210V
00
1.05pA
0
10.5pA
105pA
00
1.05nA
0
10.5nA
105nA
00
1.05µA
0
10.5µA
105µA
00
1.05mA
0
10.5mA
105mA
1
1pA-100nA ranges with PreAmp.
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......