SCPI Command Reference
17-93
Following these general steps to program and use NPLC caching:
1. Send this command to disable auto zero: SYST:AZER OFF.
2. Enable NPLC caching by sending: SYST:AZER:CACH ON.
3. Set up and run your source memory sweep with the :SOUR:MEM com-
mands along with the various other commands required to program
additional operating modes. (See
Configure memory sweep
as well as
other pertinent command descriptions in this section for details.)
NOTE
Auto zero should be disabled with the :SYST:AZER OFF command for maximum
source memory sweep speed; otherwise, the cache is of little use. With auto zero
enabled, new A/D reference and zero values are taken for every reading and saved
into the cache, slowing down sweep operation. However, with auto zero disabled,
measurements may drift and become erroneous. To minimize drift when using NPLC
caching with auto zero disabled, periodically send :SYST:AZER ONCE to force an
immediate auto zero update.
Select power line frequency setting
:LFRequency <freq>
:SYSTem:LFRequency <freq>
Select line frequency
Parameters
<freq> =
50
50Hz or 400Hz setting
60
60Hz setting
Query
:LFRequency?
Query line frequency selection.
Description
Use this command to manually select the line frequency setting (50 or
60Hz). For 400Hz operation, select 50Hz.
:AUTO <b>
:SYSTem:LFRequency:AUTO <b>
Control auto line frequency selection
Parameters
<b> =
1 or ON
Enable and line frequency selection
0 or OFF
Disable auto line frequency selection
Query
:AUTO?
Query state of auto line frequency selection.
Description
This command is used to enable or disable auto line frequency detection.
When enabled, the SourceMeter will sense the line frequency on power-up
and select the appropriate line frequency setting.
Manually setting the line frequency disables auto frequency. See the previ-
ous command.
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......