SCPI Command Reference
17-65
Select function mode
[:MODE] <name>
:SOURce[1]:FUNCtion[:MODE] <name>
Select source mode
Parameters
<name> =
VOLTage
Select voltage mode
CURRent
Select current mode
MEMory
Select memory mode
Query
[:MODE]?
Query selected source
Description
This command is used to select the source mode. With VOLTage selected,
the V-Source will be used, and with CURRent selected, the I-Source will be
used.
With MEMory selected, a memory sweep can be performed. Operating set-
ups (up to 100) saved in memory can be sequentially recalled. This allows
multiple source/measure functions to be used in a sweep.
Select sourcing mode
:MODE <name>
:SOURce[1]:CURRent:MODE <name>
Select DC sourcing mode for I-Source
:SOURce[1]:VOLTage:MODE <name>
Select DC sourcing mode for V-Source
Parameters
<name> =
FIXed
Select
fi
xed sourcing mode
LIST
Select list sourcing mode
SWEep
Select sweep sourcing mode
Query
:MODE?
Query DC sourcing mode
Description
This command is used to select the DC sourcing mode for the speci
fi
ed
source. The three modes are explained as follows:
FIXed
—
In this DC sourcing mode, the speci
fi
ed source will output a
fi
xed
level. Use the :RANGe and :AMPLitude commands to specify the
fi
xed
source level. See
Select range
and
Set amplitude for fixed source
.
LIST
—
In this mode, the source will output levels that are speci
fi
ed in a
list. See
Configure list
for commands to de
fi
ne and control the execution of
the list.
SWEep
—
In this mode, the source will perform a voltage, current or mem-
ory sweep. See
Configure voltage and current sweeps
and
Configure mem-
ory sweep
for commands to de
fi
ne the sweep.
NOTE
The sourcing mode will default to FIXed whenever the SourceMeter goes to the local
state.
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......