8-2
Data Store
Data store overview
The SourceMeter has a data store (buffer) to store from 1 to 2500 source-measure readings.
The instrument stores the source-measure readings that are displayed during the storage pro-
cess. Each source-measure reading also includes the buffer location number and a timestamp.
“Cmpl” will flash in buffer recall if the reading is in compliance.
The data store also provides statistical data on the measured readings stored in the buffer.
These include minimum, maximum, mean, and standard deviation.
NOTE
For a sweep that has a finite sweep count, the readings are automatically stored in
the buffer.
Front panel data store
Storing readings
Perform the following steps to store readings:
1.
Set up the SourceMeter for the desired configuration.
2.
Press the STORE key.
3.
Use the left and right cursor keys, and the SOURCE (
▲
and
▼
) or RANGE (
▲
and
▼
)
keys to specify the number of readings to store in the buffer.
4.
Press ENTER. The asterisk (*) annunciator turns on to indicate data storage operation.
It will turn off when the storage is finished.
5.
Turn on the output and (if necessary) trigger the unit to begin taking and storing
readings.
Recalling readings
Readings stored in the buffer are displayed by pressing the RECALL key. The source-
measure readings are positioned at the left side of the display, while the buffer location number
and timestamp are positioned at the right side.
Buffer location number
The buffer location number indicates the memory location of the source-measure reading.
Location #0000 indicates that the displayed source-measure reading is stored at the first mem-
ory location. If limit testing was performed, a “P” or an “F” will precede the buffer location
number to indicate the pass/fail result of the test. Limit testing is covered in Section 11.
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......