7-8
Relative and Math
Remote math operations
Math commands
Table 7-3 summarizes commands to control the math functions. See Section 17 for more
detailed information on these and other math commands.
Math programming example
Table 7-4 summarizes the basic command sequence for voltage coefficient testing, which is
a change in resistance of resistive elements with applied voltage. Although such changes in
resistance with voltage are present in virtually all resistors to at least some degree, voltage
coefficients are most noticeable in high-value resistors (>10
10
Ω
).
See Figure 7-2 for DUT
connections.
This example sets up the SourceMeter as follows:
•
Source function: volts
•
Sense functions: all
•
Source delay: 1sec
•
Start voltage: 10V
•
Stop voltage: 50V
•
Math expression: voltage coefficient
Table 7-3
Math commands
Command
Description
:CALCulate:MATH:NAME <name>
:CALCulate:STATe <state>
:CALCulate:DATA?
Select match expression (name = “POWER”, “OFF-
COMPOHM”, “VOLTCOEF”, “VARALPHA”).
Enable/disable math (state = ON or OFF).
Query math data.
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......