5-26
Source-Measure Concepts
Data
fl
ow
Data flow for front panel operation is summarized by the block diagrams provided in
Figure 5-15. Note that if REL is enabled, the result of the rel operation is sent to the other
blocks.
NOTE
See Appendix C for remote operation data flow information.
With Math (FCTN) and Limit Tests (LIMITS) disabled (Figure 5-15A), the SourceMeter
displays the measurement readings. If the data store is used, these readings are also stored in
the buffer for later recall. Statistical data for these readings is also available upon recall.
Figure 5-15B shows data flow when Math or Limit Tests is enabled. If Math is enabled, the
result of the math operation is displayed. If Limit Tests is enabled, the raw reading along with
the results of the tests (pass or fail) is displayed. As in the previous case, these readings can also
be stored in the data store.
Figure 5-15C shows data flow when both Math and Limit Tests are enabled. The Math oper-
ation is performed first, and then limit tests are performed on that math result. The result of the
math operation and the result of the limit tests (pass or fail) are displayed. As shown, these
readings can also be stored in the data store.
Summary of Contents for 6430
Page 26: ......
Page 32: ......
Page 78: ...2 14 Connections ...
Page 98: ...3 20 Basic Source Measure Operation ...
Page 138: ...5 30 Source Measure Concepts ...
Page 156: ...6 18 Range Digits Speed and Filters ...
Page 168: ...7 12 Relative and Math ...
Page 176: ...8 8 Data Store ...
Page 202: ...9 26 Sweep Operation ...
Page 248: ...11 22 Limit Testing ...
Page 310: ...16 6 SCPI Signal Oriented Measurement Commands ...
Page 418: ...17 108 SCPI Command Reference ...
Page 450: ...18 32 Performance Verification ...
Page 477: ...A Specifications ...
Page 489: ...B StatusandErrorMessages ...
Page 498: ...B 10 Status and Error Messages ...
Page 499: ...C DataFlow ...
Page 503: ...D IEEE 488BusOverview ...
Page 518: ...D 16 IEEE 488 Bus Overview ...
Page 519: ...E IEEE 488andSCPI ConformanceInformation ...
Page 523: ...F MeasurementConsiderations ...
Page 539: ...G GPIB488 1Protocol ...
Page 557: ......