17-24 Calibration
Model 6220/6221 Reference Manual
Return to
Calibration dates, count, and password
Viewing calibration dates and count
Front panel calibration dates and count
To view last calibration date, calibration due date, and count:
1.
Press the MENU key.
2.
Select CAL, and press ENTER. The main calibration menu will be dis-
played.
3.
Select VIEW-DATES, and press ENTER. The instrument will display the
next calibration date, last calibration date, and the calibration count. For
example:
NEXT CAL: 05/15/2005
Last cal: 05/15/2004 Count: 0001
Remote calibration dates and count
To view calibration dates and count via remote, send these queries:
For the dates, the instrument will return a string of characters with the year, day,
and month delimited by commas.
Changing the calibration password
To change the calibration password, first unlock calibration using the present
password, then enter or send the new password.
Front panel password
To change the password:
1.
Press the MENU key.
2.
Select CAL, and press ENTER. The main calibration menu will be dis-
played.
3.
Select UNLOCK, then press ENTER. Enter the present password at the
prompt, then press ENTER to unlock calibration.
CAL:PROT:DATE?
‘ Query last cal date.
CAL:PROT:NDUE?
‘ Query next cal date.
CAL:PROT:COUN?
‘ Query cal count.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com
Summary of Contents for 6220
Page 16: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 24: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 453: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...