Model 6220/6221 Reference Manual
Delta, Pulse Delta, and Differential Conductance
5-25
Return to
Communications, setup commands, and sweep table
•
The Model 622x performs a communications test and sends setup com-
mands to the Model 2182/2182A. These operations are explained in
munications test and setup commands,” on page 5-12
•
The Model 622x builds a sweep table of source values to be used for the
Differential Conductance test. For details, see
Model 2182/2182A measurement rate
The speed (rate) setting of the Model 2182/2182A is queried. If the NPLC speed
setting is not an integer value, it will be changed to 1 PLC (examples of non-integer
values are 0.1, 1.3, and 17.5).
Armed message
•
The following message is displayed briefly when Delta is armed and ready
to run:
DELTA ARMED
Press TRIG to start
Aborting Delta
After Delta is armed (or running), it can be aborted by pressing the
EXIT
key or
sending the SOUR:SWE:ABOR command.
Triggering sequence
The Trigger Link is used to synchronize source-measure triggering operations of
the Models 622x and 2182/2182A. See
“System connections,” on page 5-6
for
details on Trigger Link connections.
The triggering sequence for the first Delta cycle is shown in
and is
explained as follows:
•
Model 622x
– When Delta is started from the Model 622x, it outputs the
high (I-High) current level. After its Delta Delay expires, an output trigger
pulse is sent to the Model 2182/2182A to start its operations.
•
Model 2182/2182A
– After its Trigger Delay expires, the Model 2182/2182A
performs a measurement conversion (A/D #1). An output trigger pulse is
then sent back to the Model 622x.
•
Model 622x
– The Model 622x outputs the programmed low (I-Low) current
level. After the Delta Delay expires, an output trigger is sent back to the
Model 2182/2182A. The Delta Delay is used to allow the current source to
settle after a polarity change.
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.com
Summary of Contents for 6220
Page 16: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 24: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...
Page 453: ...Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 TestEquipmentDepot com...