Model 6220/6221 User’s Manual
Delta, Pulse Delta, and Differential Conductance
5-19
Return to
6.
Read Delta readings
– While Delta is running, the latest Model 2182/2182A
Delta reading can be read by the Model 622x using the following command:
The above read command reads the last Delta reading that was performed
by the Model 2182/2182A. If this command is sent before a new reading is
available, the last Delta reading will again be returned.
7.
When finished with Delta, it can be disarmed by sending the following com-
mand:
8.
Recall stored Delta readings
– Model 2182A Delta readings were sent to
the buffer of the Model 622x. Send the following read command to read the
buffer:
Setup and arm commands
Commands to set up and arm Delta are listed in
.
SENS:DATA?
‘ Reads the latest Delta reading.
NOTE If a read command is sent when Delta is not running,
error -221 Settings Conflict will occur.
SOUR:SWE:ABOR
‘
‘
Stops Delta and places the Model 2182A in the
local mode.
TRACe:DATA?
‘ Read Delta readings stored in 622x buffer.