Model 4200A-SCS Parameter Analyzer Reference Manual
Section 9: Keithley External Control Interface (KXCI)
4200A-901-01 Rev. C / February 2017
9-25
VL and IL
This command sets up a list sweep.
Usage
AAB
,
C
, ±
DDD.DDDD
, ±
EE.EEEE
, ... ±
EE.EEEE
AA
The source mode:
•
Voltage source (SMU or VS1...VS9):
VL
•
Current source (SMU only):
IL
B
Channel number: 1 to 9
C
Master or subordinate mode:
•
Subordinate mode:
0
•
Master mode:
1
DDD.DDDD
Compliance value:
•
Voltage source:
−
210.00 V to +210.00 V
•
Current source, 4200-SMU:
−
0.1050 A to +0.1050 A
•
Current source, 4210-SMU:
−
1.0500 A to +1.0500 A
EE.EEEE
List values:
•
Voltage source:
−
210.00 V to +210.00 V
•
Current source, 4200-SMU:
−
0.1050 A to +0.1050 A
•
Current source, 4210-SMU:
−
1.0500 A to +1.0500 A
Details
When list sweep is a selected source function, it does a list sweep with arbitrary sweep points in order
of the defined sweep parameters. If the source that is being used is a SMU, the source mode for the
sweep can be voltage or current. If, however, a voltage source (
VS1...Sn
) is being used, the source
mode must be voltage.
With the voltage source mode (
VR
) specified in the command string, you are setting the current
compliance. For the current source mode (
IR
), you are setting the voltage compliance. When
sourcing voltage, note that if you specify a compliance current that is below the minimum allowable
value, 100 pA with a preamplifier installed and 100 nA without a preamplifier, KXCI sets it to the
minimum allowable value.
The maximum number of points for the list is 4096. Sweep points must be delimited by commas.
Example
VL1,1, 0.01, 1, 5, 2
This command string sets up a channel 1 list sweep (1 V, 5 V, 2 V arbitrary steps, and compliance =
10 mA).
Also see
None