Section 6: Clarius
Model 4200A-SCS Parameter Analyzer Reference Manual
6-154
4200A-901-01 Rev. C / February 2017
Set up Cycle Mode
In the Cycle Mode, the subsite test is repeated a specified number of times. There are only measure
cycles with no stressing. For each individual test in the subsite, data is acquired for each subsite
cycle. For example, if the subsite is cycled five times, there are five sets of data and graphs for each
test. You can execute up to 128 cycles.
To set up cycle mode:
1. In the project tree, select the subsite.
2. Select
Configure
.
3. In the right pane, select the
Stress Mode Setup
tab.
4. Select
Enable Stress/Measure/Cycles
.
5. Select
Cycle Mode
.
6. Enter the
Number of Cycles
. This is the fixed number of times that you want the subsite to
execute.
7. Enter the
Cycle Delay
in seconds.
8. Select
Save
.
Figure 314: Stress Mode Setup pane
Set up stress/measure mode
To set up stress/measure mode testing, you need to:
•
Set up stress/measure mode timing
(on page 6-159)
•
Set the stress/measure mode subsite Stress Properties
(on page 6-163)
If your project is set up to run on more than one subsite, you need to set the stress properties for
each subsite separately. This allows you to have different levels of stress on each subsite. After you
configure the first site, repeat the steps for the next subsite. You can copy a subsite to speed up the
configuration process. See
Clear, copy, and paste to all sites
(on page 6-169) for instructions.