Section 6: Clarius
Model 4200A-SCS Parameter Analyzer Reference Manual
6-132
4200A-901-01 Rev. C / February 2017
Figure 292: SegARBConfig uses SeqList, SeqStartSeg, and SeqStopSeq to store sequence
In this example for the user module
PMU_SegArb_ExampleFull
, several parameters are common
across the two channels in the test:
•
SegTime (SegTime)
•
SegTrig (SegTrigOut)
•
SegMeas (MeasEnabled)
•
SegMeasType (MeasType)
•
SegMeasStart (SegMeasStart)
•
SegMeasStop (SegMeasStop)
•
SeqList (SeqList)
•
SeqStartSeg (SeqStartSeg)
•
SeqStopSeg (SeqStopSeg)
Assigning SegARBConfig to a group
As with other Control Types, to appear in the UTM Key Parameters pane, the
SegARBConfig
must
be assigned to a group. If a user module has more than one Segment Arb
®
waveform channel, place
a
SegARBConfig
control for each channel in a separate group. Choose which channel is configured
by selecting the specific variable name.
The figures below show the
SegARBConfig
for user module channel 1 and channel 2. For this
example, selecting the variable named
StartVCH1
configures the
SegStartV
for channel 1, where
selecting the variable named
StartVCH2
configures the
SegStartV
for channel 2.