Section 6: Clarius
Model 4200A-SCS Parameter Analyzer Reference Manual
6-88
4200A-901-01 Rev. C / February 2017
To select output values:
1. In the project tree, select the test.
2. Select
Configure
.
3. In the right pane, select
Test Settings
.
4. Select
Output Values
. The Output Values dialog box is displayed.
5. Select the data series that you want to send to the Analyze sheet.
6. Select
OK
.
Figure 258: Select Output Values
Compliance exit condition options
These options are available from the Test Settings pane when you select
Exit Condition
.
When the source goes into compliance, you can choose the action that the system takes.
You can select one of the following exit condition options:
•
None
: The run continues. This is the default setting.
•
Test
: The 4200A-SCS exits the test that is presently being run. If there are additional tests,
operation continues to the next test.
•
Device
: The 4200A-SCS exits the device that is presently being run. If there are additional
devices, operation continues to the next device.
•
Subsite
: The 4200A-SCS exits the subsite presently being run. If there are additional subsites,
operation continues to the next subsite.
•
Site
: The 4200A-SCS exits the site. If there are multiple sites, operation continues to the next
site.
•
Project
: The 4200A-SCS exits the project. All testing stops.
Set the compliance exit condition
To set the exit condition when compliance occurs:
1. Select the test in the project tree.
2. Select
Configure
.
3. Select
Test Settings
.
4. Select
Exit Condition
.
5. Select the option for your test and select
OK
.